LABORATORY 2 OF MEDICAL INSTRUMENTATION


Salam & Greetings !


This is the Laboratory 2 of medical instrumentation. It is a continuation from previous lab so the same circuit was used. Below are the components used for this lab.






Below are the task student need to complete. 







Student are required to select a best option of optic sensor based on the application that has been assigned to each group. For my group, we prefer to use photodiode in order to measure the level  of liquid or water.



Amplification Unit


Student required to amplify the signal of circuit from previous lab based on gain given. This means the circuit prom previous lab will be our input signal for amplifier. As the input signal in positive value, we choose to use non - inverting amplifier in order to get 10 as value of the gain. 



Image result for non inverting amplifier


This is the calculation used to get the value of resistor in order to get 10 as value of the gain.



So we used 18kOhms as R1 and 2kOhms as R2.





Circuit simulation for amplification unit





Value of gain from simulation







Circuit construction for amplification unit





Filtering Unit



Student need to choose appropriate filtering circuit in order to suppress the unwanted noise. Thus, we prefer to use low pass filter circuit. 





Low pass filter circuit




Then, student need to calculate value of capacitor they need to use in filter circuit.



C= 1/2ΠRF
=1/2Π (1.8k) (8.33 kHz)
=10.61nF




Thus, we used 10nF capacitor.





DAQ Acquisition



Student need to use LabVIEW and DAQ module in order to complete these task. 





Circuit construction using DAQ module




Block diagram in LabVIEW




Amplitude of signal



Then, student required to try 3 time for sensitivity and accuracy of circuit.


1st try

2nd try

3rd time



That conclude our laboratory 2. Thank you !